Beitrag

Miniature high-frequency chilled-mirror hygrometer for atmospheric measurements aboard fixed wing UAS

Mauz, Moritz; van Kesteren, Bram; Junkermann, Wolfgang; zum Berge, Kjell; Schön, Martin; Platis, Andreas; Bange, Jens

Bild der ersten Seite der Arbeit:

Meteorologische Zeitschrift Vol. 29 No. 6 (2020), p. 439 - 449

42 Literaturangaben

veröffentlicht: Nov 25, 2020
Online veröffentlicht: Jul 17, 2020
Manuskript akzeptiert: Apr 30, 2020
finale Ms. Revision erhalten: Apr 30, 2020
Manuskript-Revision angefordert: Apr 13, 2020
Manuskript erhalten: Jan 2, 2020

DOI: 10.1127/metz/2020/1026

BibTeX Datei

O

Open Access (Arbeit kann kostenlos heruntergeladen werden)

Arbeit kostenlos herunterladen

Abstract

A small light-weight in-house made miniature chilled-mirror hygrometer (CMH) for fixed wing UAS (unmanned aircraft system) is presented, with its features and limitations. Therefore, first measurements of the CMH equipped on the small research UAS of type MASC‑3 (multi-purpose airborne sensor carrier) operated by the University of Tübingen are shown. A comparison against a very accurate state of the art capacitive industrial humidity sensor (SHT31) is done. The sensor consists of a TEC (thermoelectric cooler) covered by a gold mirror. The TEC is controlled by a commercially available microprocessor with an on-board PID (proportional-integral-derivative) controller. The results of the CMH measurements are in good agreement with the industrial-made capacitive sensor. The absolute accuracy of the measured dew point temperature by the CMH is in the range of ±0.2 K. Spectra show evidence that the CMH is capable to measure turbulent humidity fluctuations in the atmosphere with a temporal resolution of up to 10 Hz. Such a fast humidity sensor aboard a small UAS has the potential to study humidity fluxes in the surface layer over complex terrain, behind wind energy converters and humidity variations over land and sea surfaces in general.

Schlagworte

Chilled mirror hygrometer • Unmanned Aircraft System • high frequency humidity measurement • Boundary Layer Measurements